Overview
ETCR4300 Three-phase Digital Phase Voltammeter is a multifunctional, digital and intelligent instrument specially developed by our company for field testing. It has the characteristics of high accuracy, high stability, low power consumption and convenient to use etc. Under the condition of the tested circuit does not open-loop, it can measurement of three-phase AC voltage, current, phase between voltages, phase between current, phase between current and voltage, frequency, phase sequence, active power, reactive power, apparent power, power factor, current vector sum, judge transformer wiring group, sensibility and capacitive circuit, test the secondary circuit and differential bus protection system, read the phase relationship between differential protection between each groups CT, check watt hour meter wiring correctly or not, maintenance line equipment, etc., to provide a safe, accurate and convenient new power meter for electricity inspectors.
ETCR4300 Three-phase Digital Phase Voltammeter with shake proof, skid proof, and high insulation sheath. Adopts 2.8 inch color LCD screen, dynamic display, vector diagram indication, clear and obvious, which shows the exquisite and luxurious appearance. Its sharp and small shape current clamp is suitable for the place where the wires arranging are dense.
ETCR4300 Three-phase Digital Phase Voltammeter also name intelligent three phase digital phase voltammeter, multifunction three phase digital phase voltmeter and three clamp digital phase voltmeter etc. Suitable for electric power, petrochemical, metallurgy, railway, industrial and mining enterprises, scientific research institutions, measurement departments. It is especially suitable for electric energy billing system and relay protection system.

Specification
Working Conditions
Influence Quantity
|
Base Condition
|
Working Conditions
|
Remark
|
Environment Temperature
|
23℃±1℃
|
-10℃~40℃
|
|
Environment Humidity
|
40%~60%
|
<80%
|
|
Signal Waveform
|
Sine wave
|
Sine wave
|
β=0.01
|
Signal Frequency
|
50HZ±1HZ
|
45HZ~65HZ
|
|
Meter Working Voltage
|
9V±0.1V
|
9V±1V
|
|
Current Amplitude Of Phase Frequency & Sequence Detect
|
200mA±3mA
|
2mA~20A
|
----
|
VoltageAmplitude Of Phase Frequency & Sequence Detect
|
100V±10V
|
10V~600V
|
----
|
Current Amplitude Of Power Factor detect
|
200mA±3mA
|
20mA~20A
|
----
|
VoltageAmplitude Of Power Factor detect
|
100V±10V
|
10V~600V
|
----
|
External Electric & Magnetic Field
|
To be avoided
|
The Tested Wire Position
|
Measured wire at approximately the geometric center of the clamp
|
Technical Parameters
Function
|
Description
|
CT Size
|
Φ8mm
|
Power Supply
|
DC9V alkaline batteries(1.5V AA×6)
|
Working Current
|
Max power consumption of open backlight: 135mA, battery power can continues working 6 hours
|
Meter power consumption of close backlight:90mA,battery power can continues working 9 hours
|
Display Mode
|
2.8 inch TFT color LCD screen, display area 58mmX44mm
|
Meter Size
|
196mm×92mm×54mm
|
Voltage Range
|
AC 0.00V~600V
|
Current Range
|
AC 0.0mA~20.0A
|
Phase Range
|
0.0°~360.0°
|
Frequency Range
|
45.00Hz~65.00Hz
|
Active Power Range
|
0.0W~12kW
|
Reactive Power Range
|
0.0var~12kvar
|
Apparent Power Range
|
0.0VA~12kVA
|
Power Factor Range
|
-1~+1
|
Current Vector Sum
|
0mA~60.0A
|
Resolution
|
Voltage:AC 0.01V
|
Current:AC 0.1mA
|
Phase:0.1°
|
Frequency:0.01Hz
|
Active power:0.1W
|
Reactive power:0.1var
|
Apparent power:0.1VA
|
Power factor:0.001
|
Current vector Sum:1mA
|
Phase Sequence
|
Positive phase: U1, U2, U3 or I1, I2, I3 cursors flash from left to right
Invert phase: U1, U2, U3 or I1, I2, I3 cursor flash from right to left
|
Sampling Rate
|
About2 times/second
|
Data Hold
|
PressHOLD key to keep data, "DH" symbol display
|
Data Storage
|
500 groups
|
Communication Interface
|
USB interface, store data can be upload to computer, convenient to analyze management data
|
Auto Shutdown
|
15 minutes after boot up, the meter shuts down automatically without any operation
|
Backlight Function
|
Yes, Suitable for dark place and at night use
|
Voltage Detection
|
When the battery voltage is lower than 7.2V, low battery symbol appears, reminding replace battery
|
Weight
|
Host: 550g
|
Current Clamp: 170g*3
|
Test wires: 250g
|
Test Wire Length
|
1.5m
|
Current Clamp Length
|
2m
|
Working Temperature & Humidity
|
-10℃~40℃;below 80%Rh
|
Storage Temperature & Humidity
|
-10℃~60℃;below 70%Rh
|
Input Impedance
|
Test voltage input impedance: 1MΩ
|
Withstand Voltage
|
It took 1 minute to withstand 1000V/50Hz sine wave AC voltage between the meter line and the out shell
|
Insulation
|
≥10MΩ between meter lines and sheath shell
|
Structure
|
Double insulation, with insulation vibration-proof sheath
|
Suitable Safety Standard
|
IEC61010-1 CATⅢ 600V,IEC61010-031,IEC61326,Pollution 2
|
Error & Performance Index
Category
|
Measurement Range
|
Resolution
|
Intrinsic Error
|
Voltage
|
AC 0.00V~600V
|
0.01V
|
±(0.5%range)
|
Current
|
AC 0.0mA~20.0A
|
0.1mA
|
±(0.5%range)
|
Phase
|
0.0°~360°
|
0.1°
|
±1°
|
Active Power Range
|
0.0W~12kW
|
0.1W
|
±(1.0%range)
|
Reactive Power Range
|
0.0var~12kvar
|
0.1var
|
±(1.0%range)
|
Apparent Power Range
|
0.0VA~12kVA
|
0.1VA
|
±(1.0%range)
|
Frequency
|
45Hz~65Hz
|
0.01Hz
|
±(1.0%range)
|
Power Factor Range
|
-1~+1
|
0.001
|
±0.03
|